CXL Manual v1.06
Installing transformers in the
CXL-800 rackmount housing
The CXL-800 is a steel 19” rackmount tray of open
construction; it has a depth of 225 mm, but a minimum
rack depth of 250 mm is recommended to allow adequate
clearance for rear connectors and cables.
Up to eight CXL-40T or CXL-100T transformers may be
installed in a CXL-800. Any mixture of the two types may
also be tted, up to a total of eight transformers.
The CXL-800 is not designed for CXL-200T or CXL-400T
transformers. These types should only be tted into the
CXL-1600 rackmount housing.
Each transformer is tted to the chassis baseplate with
a single M6 bolt (one supplied with each transformer).
Identify the eight pre-drilled 6 mm holes in the baseplate
(in two rows of four); insert the bolt from the underside
of the chassis, place the transformer in position with the
bolt through the central hole, and secure in place with the
washer and locknut supplied. Before nal tightening, rotate
the toroids to ensure that the cable assemblies for all the
transformers emerge from the windings at a neat orientation
for wiring to the rear panel.
If the CXL-800 is being tted with less than its full
complement of transformers, they may be mounted in any
of the positions, as convenient. Spacing them across the
width of the baseplate to give a good mechanical balance is
generally sensible.
Fitting the connectors
Note that each of the transformer positions is aligned with
one of the rear panel connector positions.
















Each transformer has ve colour-coded wires which
terminate on a small PCB. A 2-way and a 3-way 5 mm-pitch
female connector and two M3 xing brackets are tted to
the PCB. Locate the connectors into the rectangular rear
panel slot which is in line with the transformer’s position.
Note that, looking at the rear panel, the 3-way connector
goes to the left of the slot.
Secure each PCB/connector assembly in place with EITHER
two M3 x 6 mm Pozi screws, OR one M3 x 6 mm Pozi screw
and one M3 x 25 mm hex pillar. The screws and/or pillars
should be passed through the xing holes and tightened into
the tapped holes in the xing brackets. Which of the two
xing options is used depends on the connector (from 1 to
8) being tted. The diagram below shows a rear panel with
all eight connectors tted; the arrows indicate the correct
locations for the hex pillars.
Tr ansformer
Connectors
IMPORTANT: If less than eight transformers are installed
in the rackmount housing, it is still important to t all
six hex pillars (supplied) in the positions indicated. If a
connector position requiring a hex pillar is empty (due there
being no transformer in that position), use an M3 shakeproof
washer and nut (supplied) to secure the hex pillar in place.
Re-assembling the housing
Tidy the transformer wires inside the housing so that
they follow the base of the chassis. Fit the breglass safety
cover (Part No. PC340144) inside the rear panel so that
the transformer wires pass under it and it covers the rear
connectors. Note the four xing holes (arrowed) should be
uppermost. Screw the cover to the folded rear panel ange
with four of M3 x 6 mm Pozi screws.